FEATURE
Revolutionizing Oral Scanning
with High-Precision OCT Technology
Huvitz's Optical Coherence Tomography (OCT) technology is an optical imaging technique that analyzes internal structures with high precision by irradiating light. By projecting infrared light, which is harmless to the human body, it provides high resolution below 60μm. This allows for high-resolution observation of the microstructure of teeth and soft tissues. The technology enables precise imaging of deep tissues in the teeth and gums, including dental caries (cavities), microcracks, periodontal diseases, and subgingival structures.
Currently, the Cone Beam Computed Tomography (CBCT) technology used in dental applications employs X-rays and provides a resolution level of 80μm–400μm. However, CBCT has limitations in observing microstructures such as fine cracks due to its low resolution for soft tissues (e.g., gingiva and pulp). Additionally, repeated imaging poses radiation exposure risks, raising patient safety concerns.
Huvitz's OCT technology offers a non-invasive diagnostic experience, free from skin incisions or radiation exposure, ensuring the safest and most comfortable diagnostic process for patients.

Patent: TOMOGRAPHY CONVERGENCE-TYPE ORAL SCANNER
The "Tomography Convergence-Type Oral Scanner" patent describes a technology that simultaneously scans the external surface and internal structure of teeth to generate a comprehensive 3D image. By integrating Optical Coherence Tomography (OCT) scanning with intraoral scanning functions, this technology enables precise and extensive dental diagnostics. The data obtained from both scanning processes are combined through a dichroic mirror, allowing simultaneous observation of the external and internal structures of the teeth. This technology supports real-time 3D analysis, providing fast and accurate diagnostic results immediately after scanning.
[Patent No. 10-2458985, 12011336]

Enhanced Accuracy and Clarity with OCT
The Optical Coherence Tomography (OCT) interferometry imaging function penetrates gingival epithelial tissue without being affected by saliva or blood to accurately capture subgingival margins area. Experience a difference in technology with precise margin data unlike anything you've encountered before.

Simplifying Prosthesis Fabrication with AI Technology
Real-time scan images integrated automatically with margin data detected by AI streamline the process. This integration facilitates the creation of optimal prostheses through a simple verification step, followed by processing with L-CAD software, eliminating complex editing.

Faster Scanning with a Wider and Deeper Area
Designed with a wide 15x18mm Field of View (FOV) and a maximum 22mm Depth of Field (DOF), the scanner enables broader area coverage with a single movement. This ensures a rapid and smooth impression acquisition process.
Veneer mode
Cross-sectional image
of a tooth by OCTIX
Veneer Scan
Process
Pre-operation
Pre-operation
After preparation
Enamel thickness
Sub-gingival finish line
Evaluating veneers
· Analyze enamel thickness and prosthetic adhesion.
Accuracy of measured enamel layer
OCTIX
Micro CT
Micro CT
Before Prep.
3D
OCTIX
Micro CT
After Prep.
3D
OCTIX
Micro CT
Before Preparation deviation(OCT - Micro CT) : 22.6 μm
After Preparation deviation(OCT - Micro CT) : 13.4 μm
· High-resolution imaging at micrometer scale, exceeding Micro CT.
· Safe, repeatable imaging suitable for pregnant patients.
※This study was conducted Jointly by Seoul National University Dental Hospital and Huvitz.
Sub-gingival margin mode
Real Margin Point Contact
OCTiX Process(Without Cord)
Another company
The Margin Point is on the surface
OCTIX
The Margin Point is under the surface
· Distinguishes gum and tooth for precise margin lines
· No need for cord insertion to measure margins.
TABLE FOR OCTiX
You can easily install the instruments and adjust their height with a simple pedal operation. The elegant white and silver color scheme blends seamlessly into various spaces.
| DIMENSION | 580 x 590 x 784.5 mm |
| WEIGHT | 30kg |
| POWER supply | AC 220V-230V, 50/60Hz |
| Travel Length | 250mm |
| Speed | 22mm/sec |
| Load | Max 10kg |
| Others | Up, down movement operation pedal included |
TECHNICAL DATA
Model
LSO-100
Handpiece
Dimensions : 277 x 40 x 57 mm Weight : 270 g
Scanning
FOV : Up to 15×18 mm DOF : Up to 22 mm
Imaging(OCT)
FOV : 6 mm
Image Depth : Up to 5 mm
TIP
End Cross Section : 21.7 × 20.9 mm Sterilization : Autoclavable (150 times per tip)
Main Body
Dimensions : 490 × 363 × 103 mm Weight : 10 kg
Connectivity
* Distinguishes gum and tooth for precise margin lines